Leadcare® II Comparison with Graphite Furnace Atomic Absorption Spectrophotometry for Blood Lead Measurement in Peruvian Highlands

Jaime Rosales-Rimache; Manuel Chavez-Ruiz; Jorge Inolopú-Cucche; Jhonatan Rabanal-Sanchez; Lenin Rueda-Torres; Gloria Sanchez-Holguin
Indian Journal of Clinical Biochemistry · Vol. 38, Issue 3 · pp. 324-330 · 2023

Abstract

Peru is one of the countries with the highest lead contamination in the world. Biological monitoring has limitations due to the shortage of laboratories with validated methodologies for the measurement of blood lead, and it is necessary to use alternative methods for its measurement in high-altitude cities. We aimed to compare the blood lead levels (BLL) measured by the LeadCare II (LC) method and Graphite Furnace Atomic Absorption Spectrometry (GF-AAS). We measured the BLL of 108 children from the city of La Oroya. The mean and median BLL for GF-AAS were 10.77 ± 4.18 and 10.44 µg/dL, respectively; for the LC method, the mean was 11.71 ± 4.28 and the median was 11.60 µg/dL. We found a positive linear correlation (Rho = 0.923) between both methods. Notwithstanding, the Wilcoxon test suggests a significant difference between both methods ( ρ = 0.000). In addition, the Bland–Altman analysis indicates that there is a positive bias (0.94) in the LC method, and this method tends to overestimate the BLL. Likewise, we performed a generalized linear model to evaluate the influence of age and hemoglobin on BLL. We found that age and hemoglobin had a significant influence on BLL measured by the LC method. Finally, we used two non-parametric linear regression methods (Deming and Passing-Bablok regression) to compare the LC method with the GF-AAS. We found that these methods differ by at least a constant amount, and there would be a proportional difference between both. Although in general there is a positive linear correlation, the results of both methods differ significantly. Therefore, its use in cities located at high altitudes (higher than 2440 m.a.s.l.) would not be recommended.

Bibliographic Information

JournalIndian Journal of Clinical Biochemistry
PublisherSpringer
Publication Date2023-07-01
Publication Year2023
Volume38
Issue3
Pages324-330
Document TypeJournal Article
Print ISSN0970-1915
eISSN0974-0422
DOI10.1007/s12291-022-01050-y

Access Information

NARA Access Coverage1989-01-01~Current
Journal Homepagehttps://www.springer.com/journal/12291
Publisher PageOpen Publisher Page
Full-text access depends on NARA's subscribed coverage and institutional access.