Reliability and validity of the MX3 portable sweat sodium analyser during exercise in warm conditions

Harry A. Brown; Brad Clark; Julien D. Périard
European Journal of Applied Physiology · Vol. 124, Issue 7 · pp. 2153-2160 · 2024

Abstract

Purpose Accurately measuring sweat sodium concentration ([Na + ]) in the field is advantageous for coaches, scientists, and dieticians looking to tailor hydration strategies. The MX3 hydration testing system is a new portable analyser that uses pre-calibrated biosensors to measure sweat [Na + ]. This study aimed to assess the validity and reliability of the MX3 hydration testing system. Methods Thirty-one (11 females) recreationally active participants completed one experimental trial. During this trial, participants exercised at a self-selected pace for 45 min in a warm environment (31.5 ± 0.8 °C, 63.2 ± 1.3% relative humidity). Sweat samples were collected from three measurement sites using absorbent patches. The samples were then analysed for sweat [Na + ] using both the MX3 hydration testing system and the Horiba LAQUAtwin-NA-11. The reliability of the MX3 hydration testing system was determined following two measurements of the same sweat sample. Results The mean difference between measurements was 0.1 mmoL·L −1 (95% limits of agreement (LoA): − 9.2, 9.4). The analyser demonstrated a coefficient of variation (CV) of 5.6% and the standard error of measurement was 3.3 mmoL·L −1 . When compared to the Horiba LAQUAtwin-NA-11, there was a mean difference of − 1.7 mmoL·L −1 (95% LoA: − 0.25 $${\overline{\text{X}}}$$ X ¯ , 0.25 $${\overline{\text{X}}}$$ X ¯ ) and the CV was 9.8%. Conclusion The MX3 hydration testing system demonstrated very good single-trial reliability, moderate agreement and a very good CV relative to the Horiba LAQUAtwin-Na-11. To further validate its performance, the MX3 hydration testing system should be compared with analytical techniques known for superior reliability and validity.

Bibliographic Information

JournalEuropean Journal of Applied Physiology
PublisherSpringer
Publication Date2024-07-01
Publication Year2024
Volume124
Issue7
Pages2153-2160
Document TypeJournal Article
Print ISSN1439-6319
eISSN1439-6327
DOI10.1007/s00421-024-05447-w

Access Information

NARA Access Coverage1928-01-01~Current
Journal Homepagehttps://www.springer.com/journal/421
Publisher PageOpen Publisher Page
Full-text access depends on NARA's subscribed coverage and institutional access.